產品介紹∣Genova Plus Life Science Spectrophotometer
Search

Genova Plus Life Science Spectrophotometer
Genova Plus Life Science Spectrophotometer

【Jenway】火焰光度計、光譜儀 光譜儀

Genova Plus Life Science Spectrophotometer

簡介

Genova Plus

  • Life Science Spectrophotometer
  • 5nm spectral bandwidth
  • Absorbance range -0.3 to 2.5A
  • Pre-programmed for DNA/RNA analysis
  • Purity scan across entire wavelength range
  • Pre-programmed methods for protein analysis 
  • Standard spectrophotometer functions
  • Press to read xenon lamp
  • Icon driven software
  • Small footprint
  • Method and result saving to USB memory stick 

我要諮詢

網頁資料建構中

網頁資料建構中

This spectrophotometer is based on the highly successful 73 series spectrophotometers therefore it has all the same features including icon driven software and an improved navigation system for easy and intuitive usability.

As a life science spectrophotometer it has pre-programmed methods for the measurement of nucleic acid concentrations and protein assays plus measurement modes for determining the purity of nucleic acids and optical density (OD) for cell harvesting. 

It is pre-programmed with all the most common methods for protein analysis including Bradford, Lowry, Biuret, BCA and Direct UV methods. The OD measurement mode enables cell growth to be monitored by measuring absorbance at 600nm, 595nm or any other user selected wavelength. Performing a purity scan across the entire wavelength range from 198 to 1000nm displays any distorted peaks which enables impurities to be easily identified. 

This life science spectrophotometer can be used with any of the accessories of the 73 series, these include the electronic Peltier for accurate temperature control, perfect for use in kinetics experiments and temperature critical reactions, 8 cell automatic turret, sipper pump and a micro-cuvette holder for more precious samples, amongst others. 

As well as the dedicated life science measurement modes this instrument can also be used as a standard spectrophotometer with measurement modes for photometrics, concentration, multi-wavelength, spectrum scanning, quantitation and kinetics.

  • Life Science Spectrophotometer
  • 5nm spectral bandwidth
  • Absorbance range -0.3 to 2.5A
  • Pre-programmed for DNA/RNA analysis
  • Purity scan across entire wavelength range
  • Pre-programmed methods for protein analysis 
  • Standard spectrophotometer functions
  • Press to read xenon lamp
  • Icon driven software
  • Small footprint
  • Method and result saving to USB memory stick 

 

Parameter

Genova Plus

Wavelength Range

198 to 1000nm

Light source

Xenon lamp

Wavelength Resolution

1nm

Wavelength Accuracy

± 2nm

Wavelength Repeatability

± 0.5nm

Spectral bandwidth

5nm

Photometric Transmittance

0 to 199.9%

Photometric Absorbance

- 0.300 to 2.500A

Photometric Accuracy*

± 1%T, ± 0.01Abs at 1.000 Absorbance

Photometric Resolution

0.1%T, 0.001A

Concentration/Quantitation Range

0 to 9999

Concentration Calibration

Blank with a single standard or factor

Quantitation Calibration

Blank with up to 12 standards

Quantitation Curve fit algorithms

Quadratic, quadratic through zero, linear, linear through zero, interpolate

Kinetics Measurement time

2 to 9999 seconds

Kinetics Calibration

Blank with a single standard or factor

Kinetics Display

Graphical and concentration

Kinetics Analysis

Concentration, rate of change, initial and final absorbance/ %T

Spectrum/Purity Range

198 to 1000nm

Scan interval

selec table 1, 2 or 5nm

Spectrum/Purity Analysis

Absorbance or transmittance and peak and valleys

Multi-wavelength Data Points

Up to 4 wavelengths

Multi-wavelength Calculations

Sum, product, ratio and difference

DNA Measurement Modes

dsDNA, ssDNA, RNA, Oligonucleotides, 260/ 280, 260/ 230, Variable ratio

Multi-wavelength Plus Data Points

Up to 3 wavelengths plus optional reference wavelength

Multi-wavelength Plus Calculations

Concentration and ratio

Protein Measurement Modes

BCA, Bradford, Lowry, Biuret, Direct UV

OD 600 Range

0.00 E- 19 to 9.99 E+ 19

OD 600 Calibration

Blank with a single standard or factor

GLP Support

Current time and date, user ID, instrument lock and method lock

Method storage

312 (including pre-programmed methods)

Results memory

Limited by attached mass storage device

Removable media

USB (supplied)

Outputs

USB, Analogue, RS232, Integral printer

Size (w x d x h)

275 x 400 x 220mm

Weight

6kg