產品介紹∣Genova Nano Micro-volume Life Science Spectrophotometer view video
Search

Genova Nano Micro-volume Life Science Spectrophotometer view video
Genova Nano Micro-volume Life Science Spectrophotometer view video

【Jenway】火焰光度計、光譜儀 光譜儀

Genova Nano Micro-volume Life Science Spectrophotometer view video

簡介

Genova Nano Micro-volume

  • Micro-volume, life science and standard spectrophotometer measurement modes
  • Ideal for DNA, RNA and Protein measurements
  • Only 0.5µl sample volume required
  • Purity scan over entire wavelength range, 198 to 1000nm
  • Detects DNA concentrations as low as 2ng/µl 
  • Easy and quick to clean
  • Results are reproducible, accurate and easily obtained
  • Method and result saving to USB memory stick 

我要諮詢

網頁資料建構中

網頁資料建構中

The Genova Nano micro-volume spectrophotometer measures small sample volumes as low as 0.5µl with a high degree of accuracy, reproducibility and speed. It has the ability to measure small sample volumes, conserves precious samples, reduces the need for dilutions and eliminates the requirement for cuvettes. Cleaning is quick and simple; wiping the read heads with a microfibre cloth removes all trace of the sample, allowing faster change over between samples and therefore increasing sample throughput.

  • Micro-volume, life science and standard spectrophotometer measurement modes
  • Ideal for DNA, RNA and Protein measurements
  • Only 0.5µl sample volume required
  • Purity scan over entire wavelength range, 198 to 1000nm
  • Detects DNA concentrations as low as 2ng/µl 
  • Easy and quick to clean
  • Results are reproducible, accurate and easily obtained
  • Method and result saving to USB memory stick 

Parameter

Genova Nano

Wavelength Range

198 to 1000nm

Wavelength Accuracy

± 2nm

Spectral Bandwidth

5nm

Path Length

0.2 or 0.5mm (auto-ranging)

Absorbance Range

- 15 to 125A (10mm equivalent)

Absorbance Accuracy

± 2% at 260nm

Absorbance Precision

< 0.005A between 0 and 1A (at 260nm and 0.5mm)

Maximum Concentration

6,000 ng/µl (dsDNA) (at 0.2mm)

Detection Limit

2.0ng/µl (dsDNA) (at 0.5mm)

Measurement Time

< 6.5 seconds

Minimum sample Size

0.5µl (at 0.2mm),1.0µl (at 0.5mm)

Maximum sample size

5.0µl

DNA measurement modes

dsDNA, ssDNA, RNA, Oligonucleotides, 260/ 280, 260/ 230, Variable ratio

Protein measurement modes

Pierce 660, BCA, Bradford, Lowry, Biuret, Direct UV

Sample Pedestal Material

Quartz stainless steel

Light Source

Press to read Xenon lamp

Size (w x d x h), mm

275 x 400 x 220

Weight, kg

7.7