產品介紹∣9500 Bench DO2 Meter
Search

9500 Bench DO2 Meter
9500 Bench DO2 Meter

【Jenway】火焰光度計、光譜儀 Dissolved oxygen meters

9500 Bench DO2 Meter

簡介

9500 Bench DO2 Meter

• Results as % air saturation, % oxygen or mg/l
• Automatic zeroing and calibration
• Atmospheric pressure and salinity correction
• BOD5 measurement and calculation
• RS232 for connection to printer or PC and IrDA for cableless printing option

我要諮詢

網頁內容建置中

網頁內容建置中

The 9500 is a research meter designed for use in a number of applications including fish farming/breeding, all forms of aquaculture, environmental analysis, pollution control and effluent management. The 9500 includes powerful data logging capabilities with the ability to store 250 DO2 readings either manually, at timed intervals or on alarmed events. The 9500 also includes full support for the BOD5 day test protocol and has the ability to store up to 25 BOD data sets, each with up to 10 samples

Parameter

9500

DO2 range % air saturation

0 to 199%

DO2 range % oxygen saturation

0 to 25.0%

DO2 range concentration (mg/l)

0 to 19.99mg/l

Resolution

1%, 0.1%, 0.01mg/l

Accuracy

±2% within 10°C of calibration temperature

Temperature range

–10 to 60°C

Temperature resolution

0.1°C

Temperature accuracy

±0.5°C

ATC/Manual

0 to 60°C

Data Logging

250 readings storage and 20 BOD sets (10 samples per set)

Clock

24 hour, hrs/min/sec or day/month/ year, leap year corrected

GLP

Calibration reminder, alarm outputs, security code protected user data

Outputs

Analogue, RS232 and IrDA

Connector

7–pin DIN