產品介紹∣4510 Bench Conductivity Meters
Search

4510 Bench Conductivity Meters
4510 Bench Conductivity Meters

【Jenway】火焰光度計、光譜儀 Conductivity meter

4510 Bench Conductivity Meters

簡介

Conductivity Meters

• Simultaneous display of conductivity or TDS and temperature
• Automatic selection of range for best resolution
• Calibration by cell constant or standard solutions
• Automatic standard recognition
• Storage of up to 32 results
• Bi-directional RS232 link to printer or PC

我要諮詢

網頁資料建構中

網頁資料建構中

The 4510 is easy to use but with the flexibility to meet the broadest range of applications and for those where greater accuracy is required the 4510 has automatic conductivity standard recognition and endpoint detection. Set-up options include cell constant, temperature coefficient and reference temperature. The instrument can store up to 32 readings, which can also be sent to a
printer or transferred to computer via DataWay and the RS232 interface.

Parameter

4510

Conductivity range

6 auto–selected from 0 to 1999mS*

Conductivity resolution

0.01µS to 1mS

Conductivity accuracy

±0.5% ±2 digits

Automatic standard recognition

10µS, 84µS, 1413µS, 12.88mS

Cell constant

0.010 to 19.999

TDS range

6 auto-selected from 0 –1999g/l*

TDS resolution

0.01mg/l to 1g/l

TDS accuracy

±0.5% ±2 digits

EC Ratio

0.50 to 0.80

Resistivity range

NA

Resistivity resolution

NA

Resistivity accuracy

NA

Salinity range

NA

Salinity resolution

NA

Salinity accuracy

NA

Temperature range

–10 to 105°C

Temperature resolution

0.1°C

Temperature accuracy

±0.5°C

ATC and manual

0 to 100°C

Temperature coefficient

0.00% to 4.00%/°C

Reference temperature

18, 20 or 25°C

GLP

NA

Outputs

Analogue and RS232

Connector

7-pin DIN

Power

9V AC ±10% @ 50/60Hz

Size (l x w x d)

210 x 250 x 55mm

Weight

850g


* With cell constant (K)=10