Designed to measure dielectric breakdown and dielectric strength of a wide range of solid electrical insulating materials, at commercial power frequencies, within the ASTM D149 and IEC 60243-1 standards as well as other comparable international standards.
Standard output voltages up to 100 kV AC and DC
Selectable automatic or manual operating modes
Integrated test chamber with safety interlock, internal light, stress relief shielding, and transparent door
Wide variety of test fixtures with electrodes
Hot and cold oil baths
Rugged, mobile steel cabinet
Safety and Design Features
Programmable Logic Controller and Operator Terminal utilized to control all aspects of test sequences
Selectable automatic or manual operating modes
Programmable rate of rise
Over-voltage preset
Peak-reading memory voltmeter
Circuit breaker protection
Test chamber with safety interlock, stress relief shielding, and transparent door
Fast-acting as well as thermal overload protection
Over-current trip point is adjustable from 10% to 110% of rated output current
Rugged steel cabinet with durable painted finish
Casters for ease of mobility
Writing shelf on front of cabinet
Fork lifting provisions
Two copies of operation/maintenance manual
Computerized CE-Type Controls:
Auto-ranging display of voltage and current.
Accuracy of ±0.8% of reading + 0.2% of range +LSD.
Failure memory voltmeter function
Auto-voltage, over-voltage, and over-current setpoint
Stop watch, duration, dwell timer functions
Auto-step and auto-cycle functions
Motorized regulator
Variable ramp rate (desired rate of rise may be entered)
RS-232 port
Graphic display
AC units:
Input:
Voltage/Current:
6CCE20-0.8/D149: 120 V, 12 A; 230V, 7A; single phase
6CCE50-2/D149: 120V, 23A; 230V, 13A; single phase
6CCE100-5/D149: 230V, 25A; single phase
Frequency: 50 or 60Hz (one must be specified)
Output:
Voltage/Current:
6CCE20-0.8/D149: ≈0-20 kVAC, 40 mA; ≈0-3 kVAC, 100 mA
6CCE50-2/D149: ≈0-50 kVAC, 40 mA; ≈0-5 kVAC, 100 mA
6CCE100-5/D149: ≈0-100 kVAC, 50 mA; ≈0-10 kVAC, 100 mA
Duty Cycle:
Continuous: for breakdown testing
1 hr ON/1 hr OFF:
6CCE20-0.8/D149: 0.8 kVA
6CCE50-2/D149: 2.0 kVA
6CCE100-5/D149: 5.0 kVA
AC/DC units:
Input:
Voltage/Current:
6TCE4100-10/50-2/D149: 120 V, 25 A; 230V, 13A; single phase
Frequency: 50 or 60Hz (one must be specified)
Output:
Voltage/Current:
6TCE4100-10/50-2/D149: ≈0-50 kVAC, 40 mA; ≈0-100 kVDC, 10 mA
Duty Cycle:
Continuous: for breakdown testing
1 hr ON/1 hr OFF: 2.0kVA
DC units:
Input:
Voltage/Current:
4TCE40-10/D149: 120 V, 8 A; 230V, 5 A; single phase
4TCE100-10/D149: 120 V, 14 A; 230V, 8 A; single phase
Frequency: 50 or 60Hz (one must be specified)
Output:
Voltage/Current:
4TCE40-10/D149: ≈0-40 kVDC, 10 mA
4TCE100-10/D149: ≈0-100 kVDC, 10 mA
Duty Cycle:
Continuous: for breakdown testing
1 hr ON/1 hr OFF:
4TCE40-10/D149: 400 W
4TCE100-10/D149: 1000 W
Please feel free to contact for detailed specifications, accessories, software and options.