元素K對於植物生長至關重要,數十年來一直使用火焰光度法對肥料和土壤基質進行分析。BWB Soil是根據肥料和土壤基質的特定分析以及各種現場操作要求而設計。SOIL 是以屢獲殊榮的 BWB XP 型號為基礎作延伸,提供了我們基礎產品的所有優勢,並超越了農業行業的分析要求。
我要諮詢元素K對於植物生長至關重要,數十年來一直使用火焰光度法對肥料和土壤基質進行分析。BWB Soil是根據肥料和土壤基質的特定分析以及各種現場操作要求而設計。SOIL 是以屢獲殊榮的 BWB XP 型號為基礎作延伸,提供了我們基礎產品的所有優勢,並超越了農業行業的分析要求。
BWB SOIL 提供增強的鈣檢測以及鈉和鉀檢測,並可選擇利用鋰作為內部參考標準。
該儀器的便攜性和「只需添加氣體」配置允許將儀器安裝在貨車後面(需有電源供應110V以上),並在不同地理位置進行現場分析。標配熱感式印表機,可全程保持樣品分析的可追溯性。
產品特點:
Sample rate 2 – 3.5ml/min Tubing materials Silicone and Tygon®
Required desk space for Instrument 50cm (H) x 45cm (W) x 45cm (D)* For Safety reasons the Flame photometer requires 1m of unobstructed space above to allow dissipation of heat from the chimney.
Instrument size Soil Flame Photometer
Shipping
Weight Soil Flame Photometer – 14kg (30.8lbs) Shipping – 25kg (55.1lbs)
Power requirements 100V – 250V AC at 50 or 60Hz
Readout LCD, four line, alpha numeric, back lit. |
Optimal range Na 0.1 – 60ppm K 0.05 – 100ppm Ca 1.0 – 100ppm
Multi Point Calibration Na 0.1 – 1000ppm K 0.05 – 1000ppm Ca 1.0 – 1000ppm
Reproducibility <1% Coefficient of variability for 20 consecutive samples over ten minutes at concentrations of 100ppm or less. (After instrument stabilisation).
Specificity Na/K/Li = <1% to each other when equal in concentration at <100ppm
Limit of detection (LOD) and limit of quantification (LOQ) displayed LOD LOQ Na – 0.03ppm Na – 0.1ppm K – 0.02ppm K – 0.05ppm Ca – 0.3ppm Ca – 1ppm |
Time to stability
Ions measured Simultaneous measurement of Na, K, and Ca
Interfaces USB 0-2.5 volt output (based on sample concentration linked to element of users choice) Optional 4-20mA output in place of the above Integrated thermal printer .CSV and .PDF generated reports and files via FP-PC software
Recommended minimum warm up time based on ambient temperature 21ºC – 25 minutes
Fuel requirement Propane, Butane or Natural Gas* regulated to 19Bar. Flow rate of 0.4l/min *with modifications. BWB Technologies recommends either Propane or Butane for optimum results. |
Sample rate 2 – 3.5ml/min Tubing materials Silicone and Tygon®
Required desk space for Instrument 50cm (H) x 45cm (W) x 45cm (D)* For Safety reasons the Flame photometer requires 1m of unobstructed space above to allow dissipation of heat from the chimney.
Instrument size Soil Flame Photometer
Shipping
Weight Soil Flame Photometer – 14kg (30.8lbs) Shipping – 25kg (55.1lbs)
Power requirements 100V – 250V AC at 50 or 60Hz
Readout LCD, four line, alpha numeric, back lit. |
Optimal range Na 0.1 – 60ppm K 0.05 – 100ppm Ca 1.0 – 100ppm
Multi Point Calibration Na 0.1 – 1000ppm K 0.05 – 1000ppm Ca 1.0 – 1000ppm
Reproducibility <1% Coefficient of variability for 20 consecutive samples over ten minutes at concentrations of 100ppm or less. (After instrument stabilisation).
Specificity Na/K/Li = <1% to each other when equal in concentration at <100ppm
Limit of detection (LOD) and limit of quantification (LOQ) displayed LOD LOQ Na – 0.03ppm Na – 0.1ppm K – 0.02ppm K – 0.05ppm Ca – 0.3ppm Ca – 1ppm |
Time to stability
Ions measured Simultaneous measurement of Na, K, and Ca
Interfaces USB 0-2.5 volt output (based on sample concentration linked to element of users choice) Optional 4-20mA output in place of the above Integrated thermal printer .CSV and .PDF generated reports and files via FP-PC software
Recommended minimum warm up time based on ambient temperature 21ºC – 25 minutes
Fuel requirement Propane, Butane or Natural Gas* regulated to 19Bar. Flow rate of 0.4l/min *with modifications. BWB Technologies recommends either Propane or Butane for optimum results. |